Reference electrode for Aqueous solution (Ag/AgCl type) 氯化銀參考電極

參考電極選擇

電化學測量中電位的參考是基於參考電極,因此它是一個非常重要的電極。
ALS 擁有多種參考電極,包括用於水溶液、非水溶液、鹼性溶液測量的參考電極以及可逆氫電極。此外,參考電極具有標準尺寸,可以安裝在我們所有的配件中。
此外,為了完成此系列,我們還有螺旋型,可用於我們的交叉流通池徑向流通池EQCM 流通池SEC-光譜電化學流通池

以下是為您的測量樣品選擇參考電極的快速工具:

Reference electrode for Aqueous solution (Ag/AgCl type)

Reference Electrodes
RE-1B Reference electrode (Ag/AgCl)

3M NaCl is used as an internal solution.
This is the standard type of aqueous reference electrode.

RE-1B Reference electrode (Ag/AgCl)
Reference potential
AgCl+e=Ag+Cl-
E0=195 mV vs RHE (25 deg C)
∗ The reference potential is the measured value using RHEK reversible hydrogen electrode kit (Cat. No. 013597).

RE-1BP Reference electrode (Ag/AgCl)

3M NaCl is used as an internal solution.
This is the standard type of aqueous reference electrode.

RE-1B Reference electrode (Ag/AgCl)
Reference potential
AgCl+e=Ag+Cl-
E0=196 mV vs RHE (25 deg C)
∗ The reference potential is the measured value using RHEK reversible hydrogen electrode kit (Cat. No. 013597).

RE-1S Reference electrode (Ag/AgCl)

Small size for the standard type (RE-1B).

RE-1S Reference electrode (Ag/AgCl)
Reference potential
AgCl+e=Ag+Cl-
E0=195 mV vs RHE (25 deg C)
∗ The reference potential is the measured value using RHEK reversible hydrogen electrode kit (Cat. No. 013597).

RE-1CP Reference electrode (Ag/AgCl/Saturated KCl)

Aqueous reference electrode with a saturated KCl as an internal solution.

RE-1CP Reference electrode (Ag/AgCl/Saturated KCl)
Reference potential
AgCl+e=Ag+Cl-
E0=198 mV vs RHE (25 deg C)
∗ The reference potential is the measured value using RHEK reversible hydrogen electrode kit (Cat. No. 013597).

RE-3VT Reference electrode screw type (Ag/AgCl)

Aqueous reference electrode specific for ALS line-up flow cell, as Cross flow cellRadial flow cellEQCM flow cell and SEC-2F Spectroelectrochemical flow cell.

RE-3VT Reference electrode screw type (Ag/AgCl)
Reference potential
AgCl+e=Ag+Cl-
E0=195 mV vs RHE (25 deg C)
∗ The reference potential is the measured value using RHEK reversible hydrogen electrode kit (Cat. No. 013597).

For the screw type electrode, the body in PEEK material is also available.

Cat. No. Description Junction Electrolyte Isolation Purpose
012167 RE-1B Reference electrode (Ag/AgCl) IPPG 3 M NaCl Glass SVC-2, SVC-3, VC-4, Bulk electrolysis, RRDE, EQCM
013613 RE-1BP Reference electrode (Ag/AgCl) Ceramics 3 M NaCl Polymethyl pentene SVC-2, SVC-3, VC-4, Bulk electrolysis, RRDE, EQCM
013393 RE-1S Reference electrode (Ag/AgCl) IPPG 3 M NaCl Glass SECM
013691 RE-1CP Reference electrode (Ag/AgCl/Saturated KCl) Ceramics saturated KCl Polymethyl pentene SVC-2, SVC-3, VC-4, Bulk electrolysis, RRDE, EQCM
013488 RE-3VT Reference electrode screw type (Ag/AgCl) Ceramics 3 M NaCl Polymethyl pentene Flow cell (LC, EQCM, SEC-2F)

∗ IPPG: Ion Permeability Porous Glass